Human HtrA Serine Peptidase 1 (Htra1) ELISA Kit Instructions for Use

Human HtrA Serine Peptidase 1 (Htra1) ELISA Kit
  ( used in serum, plasma, cell culture supernatants and other biological fluids )
principle
This experiment used double antibody sandwich ABC-ELISA. The anti-human Htra1 mAb was coated on the microtiter plate, Htra1 in the standard and sample was combined with the mAb, and biotinylated anti-human Htra1 was added to form an immune complex attached to the plate, horseradish peroxidase The labeled Streptavidin is combined with biotin, and the substrate working solution is blue. Finally, the stop solution sulfuric acid is added, and the OD value is measured at 450 nm. The Htra1 concentration is directly proportional to the OD value, and the Htra1 concentration in the sample can be obtained by drawing a standard curve.
Kit composition ( 2-8 °C )
Coated Wells
96 holes
Enzyme Conjugate
12ml
10× specimen dilution (Sample Buffer)
12ml
20×Wash Buffer
50ml
Standards: 24ng/bottle
2 bottles
Substrate working fluid (TMB Solution)
12ml
Primary antibody working solution (Biotinylated Antibody)
12ml
Stop Solution
12ml
Prepare reagents and collect blood samples
1. Collection of specimens: serum, plasma (EDTA, heparin anticoagulation), cell culture supernatant, tissue homogenate, etc., as early as possible, stored at 2-8 ° C for 48 hours; longer time must be frozen (-20 °C or -70 °C) Save to avoid repeated freezing and thawing. 1 with sample dilution is normal before the sample measurement: 20 dilution (take 10ul, sample plus diluent 190ul, diluted 20 times).
2. Standard solution preparation: Add 0.3 ml of distilled water before use and mix well to prepare a solution of 80 ng/ml. Set 8 tubes of standard tubes, and add 300 ul of standard dilution solution to each tube. Add 80 ul of the standard solution of 80 ng/ml to the first tube, mix it, and then suck 300 ul with the sampler and transfer to the second tube. Repeat the dilution in this way, and remove 300 ul from the seventh tube and discard it. The eighth tube is a blank control.
3. The 10× specimen dilution was diluted 1:10 with distilled water (example: 1 ml concentrated dilution + 9 ml distilled water).
4. Washing solution: diluted 1:20 with distilled water (example: 1 ml concentrated washing solution added to 19 ml of distilled water)
Test procedure
1. Loading: Add 100 ul of standard or sample to be tested in each well. Mix the reaction plate thoroughly and let it stand at 37 °C for 120 minutes.
2. Wash the plate: Wash the plate thoroughly with washing solution 4-6 times, and dry it on the filter paper.
3. Add 100 ul of the first antibody working solution to each well. The reaction plate was thoroughly mixed and placed at 37 ° C for 60 minutes.
4. Wash the board: the same as before.
5. Add 100 ul of enzyme-labeled antibody working solution per well. The reaction plate was placed at 37 ° C for 30 minutes.
6. Wash the board: same as before.
7. Add 100 ul of substrate working solution per well, set 37 The reaction was carried out in the dark at °C for 15 minutes.
8. Add 100 ul of stop solution to each well and mix.
9. Measure the absorbance at 450 nm using a microplate reader within 30 minutes.
Result calculation and judgment
1. All OD values ​​should be subtracted from the blank value before calculation.
2. Using standard products 40, 20, 10, 5, 2.5, 1.25, 0.625, 0 ng/ml as the abscissa and OD as the ordinate, plot on the coordinate paper and draw the standard curve.
3. Find the corresponding Htra1 content on the graph based on the OD value of the sample, and multiply by the dilution factor.
Kit performance
1. Sensitivity: The minimum Htra1 detection concentration is less than 0.3 ng/ml.
2. Specificity: Recombinant or natural human Htra1 can be detected simultaneously. Does not cross-react with other human cytokines.
3. Repeatability: The coefficient of variation in both the plate and the plate is less than 10%.
Precautions
1. It is recommended to make double holes for the above standard holes and samples to be tested. The standard curve should be made at the same time for each measurement.
2. The washing process is critical. Insufficient washing will result in an accuracy error and an erroneous rise in the OD value.
3. After the slats are opened, the remaining slats should be sealed again to keep the slats dry .
4. This kit should be stored in a 4 o C refrigerator.
5. This kit is for scientific research only and cannot be used for clinical diagnosis!

Syringe

The function of syringe set mainly contains sealing function, tunnel permeability, tensile strength, puncture performance, degree of coordination, particulate contamination, reducing substances, metal ions, ph, ethylene oxide residue, sterility, bacterial endotoxin, cytotoxicity, sensitization, intradermal reaction, hemolysis, acute systemic toxicity, etc.


A mixing system sterile medicine mixer (powder and liquid ) is a device that mixes powder and injection before injecting the drug into the patient.But traditional booster injection infusion dispensing still exist, need medicine soluble medicine, when dispensing, repeated shocks and mixed extraction liquid, in the process of repeated puncture easy to produce plastic particles, particles stopper is getting into the human body easily blocked blood vessels, granulation, some problems of inflammation, and the ward of bacteria, viruses, repeatedly extracting liquid easy to carry the virus in the air, infusion to the solution of the problem.