BSP-100/150/250/400 liquid crystal biochemical incubator instruction manual

BSP biochemical incubator is a high-precision thermostat with cold and heat control. It can be used for plant culture, breeding test, culture, preservation of bacteria, mold, microorganisms, BOD determination of water body analysis and constant temperature test for other purposes. It is an ideal test equipment for scientific research units such as biological genetic engineering, medical treatment, health and epidemic prevention, drug testing, agriculture, animal husbandry and aquatic products.

Structural characteristics of liquid crystal biochemical incubator The chassis body is made of high quality thin plate, and the surface is painted firmly and beautifully. The left side of the door is fitted with a double-glazed glass window and a translating exterior door. When observing the experimental items in the box, just move the outer door to the right and open the lighting inside the box to make the test items clear at a glance. The inner liner is made of mirror stainless steel sheet, and the four corners are rounded, which makes the temperature inside the box uniform, durable and easy to clean. The temperature control part adopts a new intelligent digital display temperature adjustment instrument with high brightness LCD display. Users can set and adjust the temperature and time by operating the touch panel of the control panel according to different requirements to achieve the test purpose. Casters on the bottom of the cabinet make installation and movement easier and more convenient.

Working principle of liquid crystal biochemical incubator The incubator is converted into an electrical signal by the actual temperature felt by the temperature sensor located in the box, and the microcomputer or the refrigeration compressor is controlled by the microcomputer to reach the required temperature.

LCD biochemical incubator technical parameters
1. Nominal volume: 100L, 150L, ​​250L, 400L;
2, temperature control range: 5 ~ 50 ° C;
3. Temperature fluctuation tolerance: ±0.5 °C;
4, temperature uniformity tolerance: ± 1 ° C;
5, the power supply voltage: AC 220V / 50Hz;
6, input power: 1000W (100L); 1100W (150L, ​​250L); 1350W (400L);
7. Working environment: the ambient temperature is 10~30°C and the relative humidity is below 70%;
8, refrigerant: R134;
9. Equipment category: Class I.

LCD biochemical incubator installation
1. This equipment should be installed in a room that is away from direct sunlight and is ventilated and dry. The equipment and the wall must have a distance of more than 10CM.
2. The box body can be fixed in place to fix the front two brake wheels, so that the box is placed stably.
3. This equipment uses AC 220V/50Hz power supply, and there must be a reliable grounding wire in the power supply circuit to ensure safe use.

Liquid crystal biochemical incubator use method
1. Temperature controller panel layout instructions 1 SET button is used for parameter setting;
2 Move button is used to set the movement;
3 minus key is used for parameter modification;
4 plus key for parameter modification;
5 The LCD shows various types (for display status).
a) the current value shows the measured temperature or humidity;
b) The set value shows the set value temperature or humidity;
c) When the upper limit alarm is displayed, the measured temperature exceeds the upper limit;
d) indicates that the compressor starts to work when the compressor is started;
e) The temperature and humidity sensor are working properly when the control indication is displayed.
2. The temperature setting power switch is pressed to “on”, the biochemical incubator enters the working state, and the LCD screen displays various prompts and parameters.
1) Press and hold the move button first, then press the SET button to enter the temperature setting state (current value display C-01), the set value single digit flashes, press the move button to move the flashing digit, press plus or minus Press to set the desired temperature.
2) Press SET again to enter the timing setting state (current value display ┌-01), the single digit of the set value flashes, press the move button to move the flashing digit, press the plus or minus button to set the required timing. time. (Note: The time unit is minutes, and the set value is 0 for continuous operation.)
3) Press the SET button again to display the temperature inside the box. The set value shows the set value. At this point the temperature has been set and can be used.
Liquid crystal biochemical incubator debugging method
1. A method of correcting the difference between the temperature inside the box and the display temperature.
Prerequisites: The device has the function of correcting temperature. Because the equipment varies with the area, climate and usage environment, when you perform temperature detection, you may find that the temperature measurement equipment (such as thermometer) is used in the box. The temperature (called the actual temperature) does not match or differ greatly from the temperature displayed by the temperature controller (called the display temperature), and temperature correction can be performed at this time.
1. In the display state, press the “SET” button, the current value: display SEL status, when the setting value is 0000, press the move button to move the flashing digit, press the plus or minus button to enter the value “0111”.
2. Press the “SET” button again to enter the temperature parameter correction mode. The current value is displayed as: 1SC.
The value of the set value is adjusted. If the actual temperature (the temperature measured by the thermometer) is different from the displayed temperature, the measured value of the studio thermometer is subtracted from the displayed temperature value of the meter, and the result is the actual correction value of "1SC". . If the measured value of the working temperature of the working chamber is smaller than the temperature value displayed by the meter, the corrected value obtained is a negative value, and vice versa. Use the move button to move the flashing number and press the plus or minus button to set the desired temperature.
3. After setting, press and hold the “SET” button, the current value will display “1HL”, set the slope correction, do not change the default value (100), and then press the SET button continuously to return to the display state. It has been corrected to work.
4, judge the premise
a) The thermometer shall be measured after passing the inspection by the inspection department.
b) Measurement of the installation position of the mercury thermometer: The thermometer is suspended in the center of the inner cavity and cannot be measured directly on the partition.
2. Set the upper limit temperature alarm value, timing, month, day, hour, minute, print time, and ambient temperature.
1) In the display state, press the SET button, the current value displays SEL, the set value displays “0000”, press “Add” to enter the value 0001, and when the set value displays “0001”.
2) Press SET again, the current value shows AL1 as the upper temperature alarm value, then press “△” or “▽” to set the desired (factory setting is 2), then press “SET” to return to display. status.
3) Press SET again, the current value shows AL2 as the humidity upper limit alarm value, and then press “△” or “▽” to set the desired (factory setting is 5, only for constant temperature and humidity box), and then continuous Press “SET” to return to the display state.
4) Press SET again, the current value shows T as the time setting value, then press “△” or “▽” to set the required time setting value (factory setting is 0), then press “Continuously” SET" will return to the display state.
5) Press SET again, the current value shows nd as month and day, then press “△” or “▽” to modify the month and day, then press “SET” continuously to return to the display state.
6) Press SET again, the current value shows hn as hour and minute, then press “△” or “▽” to modify the hour and minute, then press “SET” continuously to return to the display state.
7) Press SET again, the current value shows Tprn as the setting time of the index interval time, then press “△” or “▽” to set the required interval time (factory setting is 60), then Press “SET” continuously to return to the display state.
8) Press SET again, the current value shows °C as the ambient temperature, then press “△” or “▽” to modify the ambient temperature value (factory setting is 26°C), then press “SET” to return to the display state.
3. When the ambient temperature changes greatly or the display temperature fluctuates greatly.
In order to ensure the temperature control accuracy of the incubator, it is necessary to start the self-tuning function of the instrument, so that the technical parameters and the temperature rise curve in the controller are adjusted to the optimal state at the ambient temperature, thereby achieving precise control of the temperature.

Note:
Self-tuning function operation
1. In the display state, press the “SET” button, the current value: display SEL status, press the plus or minus button to set the value to 0534 when the value is set to 0000.
3. Press the “SET” button again to enter the temperature parameter correction mode. The current value shows 9CI.
4. After pressing the “SET” button for nine times, the current value displays AT1 and press “Add” to enter the value 0001.
5. When the set value shows “0001”, the LCD will display “Auto-tuning” (automatically disappear after auto-tuning).
6. Press “SET” continuously to return to the display state (the temperature in the self-tuning box will fluctuate greatly).
☆Friendly reminder: In order to ensure the accuracy of the temperature inside the box, please turn off the fresh air switch when you do not need to open the fresh air!
1. When the temperature inside the box is set above 38 °C, if the temperature drifts, please check if the ambient temperature is within 30 °C. If the ambient temperature exceeds 30 °C, continue to use it. Please re-do it. Set the cabinet environment parameters (the default is 26 ° C, it has been set at the factory).
2. If the ambient temperature has exceeded 30 °C, please set the default ambient temperature of the cabinet to ≥28 °C; when the ambient temperature is below 10 °C, please set the cabinet ambient temperature to ≤25 °C.
Specific operation steps: Press “SET” in the display state, the current value: display SEL status, when setting value 0000, press the move key to move the flashing digit, press the plus or minus key to enter the value “0001” → press the setting continuously The key → ° C (26 ° C factory default) is modified.

Precautions and maintenance
1. During the handling of the incubator, it is forbidden to invert and lay flat more than 45 degrees.
2. Do not change the set value frequently during use to avoid overloading the compressor and affecting the service life of the device.
3. When there is no need for lighting in the box, the lighting switch should be placed in the "off" position to avoid affecting the upper layer temperature and prolong the life of the lamp.
4. This machine is equipped with two sets of fuses. If there is a malfunction during operation, please cut off the power supply first, check if the fuse is intact, and then check other parts.
5. In order to maintain the appearance of the device, do not wipe the surface with a corrosive solution. Wipe the cloth with a dry cloth or alcohol to keep the inside of the box clean.
6. When the equipment is not in use, keep the inside of the box dry and cut off the power.
7. To ensure uniform temperature inside the box, always check the normal operation of the axial fan in the box. During the experiment, the contents of the box should not be placed too densely and do not block the fan outlet to facilitate circulation of the airflow inside the box.
8. Do not touch, touch the temperature probe inside the box, causing the temperature to run out of control.
9. If the equipment fails, you should ask a professional to repair or contact the sales department of the factory. Users should not arbitrarily dismantle.
10. Our products implement “three guarantees”. From the date of product sale, if there are quality problems within one year, the factory is responsible for free maintenance (except for man-made damage). Exceeding the "three guarantees" period of discretionary repairs.